Crescent Healthcare
Home Contact
Sign in
  • Home
  • Product Support
  • Linet
  • Eleganza 5
  • How To | Eleganza 5 - I-Drive Power Troubleshooting

How To | Eleganza 5 - I-Drive Power Troubleshooting

This article provides step-by-step guidance to troubleshoot issues with your Eleganza 5 I-Drive Power.

    Can't find it?

    No article matches that yet. Contact our team and we'll help.

    1 min read Download PDF

    In case of a battery discharge or malfunction, the Eleganza 5 has an emergency function to ensure the patient in bed and the carer are both safe, with options to carry out bed functions manually.

    Use the emergency I-Drive Power wheel retraction, in case of emergency shutdown or if you are unable to restart the system due to battery discharge or general bed malfunction.

    To start, press any green GO button on the Eleganza 5 then press and hold the emergency retraction button located under the chassis cover on the left side of the bed. Hold the button until the fifth wheel is completely retracted.

    When the Eleganza 5 fifth wheel is retracted, it allows for manual mobility of the bed for better steering and manoeuvring making it easier for users to troubleshoot i-Drive power issues.

    Watch the video below to learn more about Eleganza 5 - I-Drive Power troubleshooting.

     

    i-drive troubleshoot eleganza 5 eleganza 5 bed i-drive power i-drive power troubleshooting idrive idrive power

    Was this article helpful?

    Yes
    No
    Give feedback about this article

    Still stuck?

    Our clinical team can help.

    Contact us

    Related articles

    • How To | Eleganza 5 - I-Drive Power
    • How To | Eleganza 5 - Side Rail Override
    • How To | Eleganza 5 - Mobi-Lift and Patient Egress
    • How To | Eleganza 5 - Trendelenburg/Reverse Trendelenburg
    © 2026 Crescent Healthcare Pty Ltd 1300 04 05 06 · [email protected]
    Expand