How To | Eleganza 4 - Scale
Learn how to use the Eleganza 4 Scale, a built-in feature that helps weigh patients without leaving the bed.
The Eleganza 4 bed enables patients to be weighed in bed without compromising comfort and accuracy.
The built-in scale can be used to weigh the patient, follow the steps below:
- Check that the patient is out of bed.
- Zero the scale and make sure all standard linen and equipment are in place.
- Press any green go button to start.
- Press and hold the “ZERO” button on the safety monitor, until a beep sound is heard and the weight change indicator flashes, to set the Eleganza 4 Scale to zero.
- Release the button, press and hold the zero button again until another beep sound is heard, with the patient on the bed.
- Allow the scale stability indicator to become solid and stop blinking to record the weight.
- Display weight in kilograms or pounds by pressing and holding the “KG/LB” button changing depending on preferences.
- Cancel zeroing the weight by pressing the “WEIGHT CLEAR” button until the weight flashes.
- Ensure to press the freeze button before adding or changing bed accessories.
- Press and hold the freeze button and wait until the icon is displayed on the screen to start adding, removing, or changing bed items.
Note: The weight stability indicator will turn solid and show when it is good to press the freeze button again to display the original weight.
To learn more about Eleganza 4 Scale, watch the video below.