Info | Reconditioning - CuroCell A4 - Instructions of Use
Gain insights on the correct and most effective use of the CuroCell A4 to ensure clean and hygienic results.
Table of Contents
Control Unit
Wipe off the control unit with a damp cloth and mild cleaning solution for this purpose: such as alcohol with or without tenside surfactants, oxidising solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %.
If using a different agent, choose one that does not affect the control unit.
Mattress
- Disconnect the tube connector from the control unit and remove the air from the mattress.
- Remove the covers. For more information of cleaning the covers, refer to the article on Cleaning - CuroCell A4 – Instructions of Use.
- Move the mattress to a clean area that is suitable for cleaning.
- Wipe off the cells, all tubes and the CPR module with a mild cleaning solution for this purpose such as: alcohol with or without tenside surfactants, Isopropanol 70 %, oxidising solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. Let it dry.
Note: When using the CuroCell® CX15 or CuroCell® CX20, the air cells can be divided using a quick‐release connector, making it easier to clean the cells. - Wipe off the work surfaces using a suitable detergent and/ or disinfectant.
- Put the mattress together. If cells for any reason have become loose from the tubes, these must be replaced according to the drawing/images under Mattress Product Description - CuroCell A4 – Instructions of Use.
Foam Core
Clean the affected area with a mild detergent (such as washing-up liquid) and water or with an alcohol-based disinfectant (cleaner intended for this purpose. Gently squeeze out any water.
NOTE:
Check the hygiene cover, cells and hoses each time the product is cleaned. If damaged, it must be replaced or repaired. Also check the control unit, tube connectors and power cable during cleaning. Damaged parts must be replaced or repaired.
Do not wring or roll the foam core to extract the water. Let it dry in a warm, ventilated area (not in direct sunlight). The foam core must be completely dry before it is used again.
For more information and to learn more about the CuroCell A4 product line, along with other Care of Sweden products, visit the Crescent Healthcare website, and check out the articles under the Care of Sweden category.