Info | Control Unit Functions - CuroCell A4 - Instructions of Use

Gain knowledge of the functions of the Control Unit for CuroCell A4 with clear instructions for use.

    Can't find it?

    No article matches that yet. Contact our team and we'll help.

    1 min read Download PDF

    On this page

    Button Function Automatic Setting Programs Maximum Pressure (Caring Mode) Panel lock Comfort Settings

    Button

    Function

    Mute the information signal
    Panel lock
    Pack & Go. Automatic deflation of the system
    Alternating mode
    Pulsating mode
    CLP mode (Constant low pressure)
    Maximum pressure (caring mode)
    Information signal
    Incorrect connection of the air connector (CPR)
    Cycle time settings (10, 15, 20, 25 minutes). The diodes are also used for error notifications.
    Comfort settings

    Automatic Setting

    The mattress system regulates the mattress's internal pressure independently and without manual adjustment to different values based on the user's weight and position. No manual action is required to affect the internal pressure of the mattress. This function works in the following three ways: 

    1. The control unit carries out an automatic setting immediately after switching the system on. 
    2. During usage, if any significant change occurs, the control unit will perform an automatic setting. 
    3. When in use, the control unit will carry out automatic settings at fixed intervals to ensure correct control of the pressure in the mattress at all times. 

    Once the automatic setting has completed, the control unit will either start in the default Pulsating mode (with first time use) or in the previously set mode function.

    Programs

    Three programs are available. Press the button to switch between the programs. We recommend the Pulsating mode as the basic setting and this is also the default setting from start.

    1
    Alternating mode. The pressure alternates between the cells. The cycle period can be changed according to the user needs and requirements. Choose between 10, 15, 20 or 25 minutes. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes.
    2
    Pulsating mode. A comfort mode that combines Constant low pressure (CLP) with Alternating mode. The cycle period can be changed according to the user needs and requirements. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes. 
    3
    CLP mode. The pressure will be the same in all cells. This is also called ‘Constant low pressure’. 

    Maximum Pressure (Caring Mode)

    With this function, the entire mattress is inflated and provides firmed support. This function automatically reverts to the previous setting after approx. 20 minutes. The function should be used when caring the user, shifting the user’s position or moving the user in or out of bed.

    Panel lock

    Press the Key Lock button to lock or unlock the control panel. The button indicates when the panel has been locked. The screen locks automatically if left untouched for five minutes. This is to prevent the settings being changed accidentally. 

    To unlock, press the button for 2 seconds.

    Comfort Settings

    Depending on the user needs, the air pressure in the mattress can be adjusted. The adjustments are done based on the automatic setting performed by the control unit (0) according to the user weight. 

    Note: When only parts of the mattress are under load, for example, in the case of amputees – it may be necessary to raise the setting using the comfort settings.

    The selected setting is shown by a green light.

    For more information and to learn more about the CuroCell A4 product line, along with other Care of Sweden products, visit the Crescent Healthcare website, and check out the articles under the Care of Sweden category.

    functions controls control unit functions instructions of use curocell a4

    Was this article helpful?

    Yes
    No
    Give feedback about this article

    Still stuck?

    Our clinical team can help.

    Contact us

    Related articles

    • Info | Quick Reference Guide - CuroCell A4
    • Info | How can I tell when the mattresses need to be replaced?
    • Info | Do I get a better mattress by choosing air instead of foam?
    • Info | What should I consider when making the bed?
    • Info | What you should consider when making the bedding for a user at risk for pressure injuries
    © 2026 Crescent Healthcare Pty Ltd 1300 04 05 06 · [email protected]
    Expand